Measuring system | Zirkonium-dioxide current probe |
Probe signal | -30 ... 150 mV |
Exhaust gas temperature | LS2 up to 300 °C; LS2-K up to 400 °C; LS2-Ex up to 500 °C LS-HT up to 1200 °C |
Heating-up time | In 10 min. ready for measuring |
Setting time after measurement value jump | t90 < 15 s |
Range of measuring | 0-21 Vol. % O2 |
Preciseness of measuring | ± 10. % of measurement value, not better than ±0,3 Vol.% |
Protection type | Optional: Ex (Zone I + II)* |
* LS2 Ex/IIB with Korean approval available
The Lambda probe LS2 consists out of an electrochemical cell, made of Zirkonium-dioxide-ceramic for direct O2 measurement of exhaust gases in hyperstoichiometric range Lambda > 1.
In connection with the Lamtec Lambda Transmitter LT2, the probe becomes a simple "plug-and-play" O2 analyzer - without having to calibrate with test gas. If higher accuracies are desired, then a calibration with test gas is possible.
The measurement is directly in the wet flue gas. The probe connection is carried out via plug, so that the handling is very easy.
Temperature | Area of use | Specials | |
LS2 (standard version) | up to 300 °C (standard version) | Heating oil EL, Gas | CO/O2 monitoring in combination with probe KS1D Outdoor usage: with rain cover |
LS2-K | up to 450 °C (calibration version) | Heating oil E/SL, Gas, Coal, Solids (Wood, Bio fuel), Special fuels on request | CO/O2 monitoring in combination with probe KS1D Suitable for outdoor usage yes: able to calibrate |
LS2-E | > 400 °C to1400 °C (ejector extraction) | Heating oil EL, Gas | CO/O2 monitoring in combination with probe KS1D Outdoor usage: Yes |
LS2-Ex | up to 500 °C (Ex zone version) | Heating oil E/SL, Gas, Coal, Solids (Wood, Bio fuel), Special fuels on request | CO/O2 monitoring in combination with probe KS1D Outdoor usage: Yes Ex design |
LS2-HT | up to 1200 °C (high temperature version) | Heating oil EL, Gas, special fuels on request | CO/O2 monitoring in combination with probe KS1D Outdoor usage: Yes |